Analysis of Intra-Aortic Balloon Performance in Open-Heart Surgery

Authors

  • Zainab A. Wajeeh Dept. of Bio-Medical Engineering, Al-Nahrain University, Baghdad-Iraq.
  • Sadiq J. Hamandi Dept. of Bio-Medical Engineering, Al-Nahrain University, Baghdad-Iraq.
  • Wisam S. Alobaidi Cardiac surgical department of Ibn Al-Bitar cardiac surgical center, Baghdad, Iraq.
  • Georges B. Tedy Cardio-Vascular Surgery, President of The Euro-Asian and Lebanese Society of Cardiac Surgery, Member of The French Society of Cardiac Surgery, Member of The Society of The Thoracic Surgeons.

DOI:

https://doi.org/10.29194/NJES.26030157

Keywords:

Intra-Aortic Balloon Pump, Open-Heart Surgery, Counter-Pulsation, Mistiming

Abstract

A failing heart can be supported in several ways, including cardiopulmonary bypass pumps (CPB), extracorporeal membrane oxygenators (ECMOs), and other types of auxiliary heart pumps. The intra-aortic-balloon-pump (IABP) is one technique of internal counter-pulsation that supports maintaining the circulatory system It continues to be used as a vascular support device to critically unwell cardiac patients. Many recent studies have focused on the problems of the (IABP) in open-heart surgery, while other researchers concentrated on the positioning and size of the balloon, some of them studied the timing of the balloon's inflation and deflation. this paper has reviewed a brief Introduction, the basic principles of the balloon, how to trigger the balloon pump as well as the use of IABP in Coronary Artery Bypass Graft (CABG), balloon mistiming of inflation and deflation, balloon timing usage within open-heart surgery and finally a balloon position and sizing.

Downloads

Download data is not yet available.

References

A. Kantrowitz, “Origins of intra-aortic balloon pumping,” Ann Thorac Surg, vol. 50, no. 4, pp. 672–674, 1990, Doi: 10.1016/0003-4975(90)90220-Z.

M. Krishna and K. Zacharowski, “Principles of intra-aortic balloon pump counterpulsation,” Continuing Education in Anaesthesia, Critical Care and Pain, vol. 9, no. 1, pp. 24–28, 2009, Doi: 10.1093/bjaceaccp/mkn051.

B. Mottahedi, J. Esfehanizadeh, K. Alizadeh, and Z. A. Shaye, “The Evaluation of Survival in Patients Who Need Intra-Aortic Balloon Pump (IABP) after Cardiac Surgery,” 2014.

T. Kishimori et al., “Intra-aortic balloon pump and survival with favorable neurological outcome after out-of-hospital cardiac arrest: A multicenter, prospective propensity score-matched study,” Resuscitation, vol. 143, pp. 165–172, Oct. 2019, Doi: 10.1016/j.resuscitation.2019.07.002

T. G. Papaioannou and C. Stefanadis, “Basic principles of the Intra-aortic balloon pump and mechanisms affecting its performance,” ASAIO Journal, vol. 51, no. 3. pp. 296–300, May 2005. Doi: 10.1097/01.MAT.0000159381.97773.9B

K. Gu et al., “Numerical analysis of aortic hemodynamics under the support of venoarterial extracorporeal membrane oxygenation and intra-aortic balloon pump,” Comput Methods Programs Biomed, vol. 182, Dec. 2019, Doi: 10.1016/j.cmpb.2019.105041

H. Thiele et al., “Intra-aortic Balloon Pump in Cardiogenic Shock Complicating Acute Myocardial Infarction: Long-Term 6-Year Outcome of the Randomized IABP-SHOCK II Trial,” Circulation, vol. 139, no. 3, pp. 395–403, Jan. 2019, Doi: 10.1161/CIRCULATIONAHA.118.038201.

V. Gramigna, M. v. Caruso, M. Rossi, G. F. Serraino, A. Renzulli, and G. Fragomeni, “A numerical analysis of the aortic blood flow pattern during pulsed cardiopulmonary bypass,” Comput Methods Biomech Biomed Engin, vol. 18, no. 14, pp. 1574–1581, Oct. 2015, Doi:10.1080/10255842.2014.930136

M. V. Caruso, V. Gramigna, and G. Fragomeni, “A CFD investigation of intra-aortic balloon pump assist ratio effects on aortic hemodynamics,” Biocybern Biomed Eng, vol. 39, no. 1, pp. 224–233, Jan. 2019, Doi: 10.1016/j.bbe.2018.11.009.

J. R. C. Jansen, M. B Bastos, P. Hanlon, N. M. van Mieghem, O. Alfieri, and J. J. Schreuder, “Determination of cardiac output from pulse pressure contour during intra-aortic balloon pumping in patients with low ejection fraction,” J Clin Monit Comput, vol. 34, no. 2, pp. 233–243, Apr. 2020, Doi:10.1007/s10877-019-00320-0

L. S. González and M. A. Chaney, “Intra-aortic balloon pump counterpulsation, part I: History, technical aspects, physiologic effects, contraindications, medical applications/outcomes,” Anesthesia and Analgesia, vol. 131, no. 3. Lippincott Williams and Wilkins, pp. 776–791, Sep. 01, 2020. Doi:10.1213/ANE.0000000000004954

J. M. Ali and Y. Abu-Omar, “The intra-aortic balloon pump and other methods of mechanical circulatory support,” Surgery (United Kingdom), vol. 36, no. 2. Elsevier Ltd, pp. 68–74, Feb. 01, 2018. Doi: 10.1016/j.mpsur.2017.11.002.

C. A. J. Webb, P. D. Weyker, and B. C. Flynn, “Management of intra-aortic balloon pumps,” Semin Cardiothorac Vasc Anesth, vol. 19, no. 2, pp. 106–121, Jun. 2015, Doi: 10.1177/1089253214555026.

A. S. Franco, A. C. Bridi, M. de A. Karam, A. P. A. Moreira, K. B. S. de Andrade, and R. C. L. da Silva, “Stimulus-response time to alarms of the intra-aortic balloon pump: safe care practices,” Rev Bras Enferm, vol. 70, no. 6, pp. 1206–1211, Nov. 2017, Doi: 10.1590/0034-7167-2016-0432.

M. Khorsandi et al., “A 20-year multicentre outcome analysis of salvage mechanical circulatory support for refractory cardiogenic shock after cardiac surgery,” J Cardiothorac Surg, vol. 11, no. 1, Nov. 2016, Doi: 10.1186/s13019-016-0545-5.

G. F. Serraino et al., “Pulsatile cardiopulmonary bypass with intra-aortic balloon pump improves organ function and reduces endothelial activation,” Circulation Journal, vol. 76, no. 5, pp. 1121–1129, 2012, Doi: 10.1253/circj.CJ-11-1027.

G. Yumun et al., “Analysis of clinical outcomes of intra-aortic balloon pump during coronary artery bypass surgery,” Cardiovasc J Afr, vol. 26, no. 3, pp. 130–133, May 2015, Doi: 10.5830/CVJA-2015-010.

M. Nowak-Machen et al., “Influence of Intra-aortic balloon pump counterpulsation on transesophageal echocardiography derived determinants of diastolic function,” PLoS One, vol. 10, no. 3, Mar. 2015, Doi: 10.1371/journal.pone.0118788.

H. Hugo Escutia-Cuevas, “Use of Intra-Aortic Balloon Pump During Coronary Artery Bypass Graft Surgery. Current Questions and Few Answers. Use of Intra-Aortic Balloon Pump During Coronary Artery Bypass Graft Surgery. Current Questions and Few Answers,” 2021, Doi: 10.22541/au.161262483.32657582/v1.

I. Ahmad, M. U. Islam, M. U. Rehman, and B. Khan, “Frequency of intra-aortic balloon pump insertion and associated factors in coronary artery bypass grafting in a tertiary care hospital,” Pak J Med Sci, vol. 37, no. 2, pp. 1–5, Mar. 2021, Doi: 10.12669/pjms.37.2.3614.

H. Parissis, A. Soo, and B. Al-Alao, “Intra-aortic balloon pump: Literature review of risk factors related to complications of the Intra-aortic balloon pump,” Journal of Cardiothoracic Surgery, vol. 6, no. 1. Nov. 02, 2011. Doi: 10.1186/1749-8090-6-147.

K. de Silva et al., “Coronary and microvascular physiology during intra-aortic balloon counterpulsation,” JACC Cardiovasc Interv, vol. 7, no. 6, pp. 631–640, 2014, Doi: 10.1016/j.jcin.2013.11.023.

A. R. Ihdayhid, S. Chopra, and J. Rankin, “Intra-aortic balloon pump: Indications, efficacy, guidelines and future directions,” Current Opinion in Cardiology, vol. 29, no. 4. Lippincott Williams and Wilkins, pp. 285–292, 2014. Doi: 10.1097/HCO.0000000000000075.

T. Ahmad et al., “Effects of left ventricular assist device support on biomarkers of cardiovascular stress, fibrosis, fluid homeostasis, inflammation, and renal injury,” JACC Heart Fail, vol. 3, no. 1, pp. 30–39, Jan. 2015, Doi: 10.1016/j.jchf.2014.06.013.

C. de Lazzari et al., “Intra-aortic balloon counterpulsation timing: A new numerical model for programming and training in the clinical environment.,” Comput Methods Programs Biomed, vol. 194, Oct. 2020, Doi: 10.1016/j.cmpb.2020.105537.

C. Kolyva, G. M. Pantalos, J. R. Pepper, and A. W. Khir, “Does conventional intra-aortic balloon pump trigger timing produce optimal hemodynamic effects in vivo?,” International Journal of Artificial Organs, vol. 38, no. 3, pp. 146–153, 2015, Doi: 10.5301/ijao.5000385.

A. Donelli et al., “Performance Of A Real-Time Dicrotic Notch Detection And Prediction Algorithm In Arrhythmic Human Aortic Pressure Signals,” 2002.

J. J. Schreuder et al., “Automatic Intra-aortic balloon pump timing using an intrabeat dicrotic notch prediction algorithm,” Annals of Thoracic Surgery, vol. 79, no. 3, pp. 1017–1022, 2005, Doi: 10.1016/j.athoracsur.2004.07.074.

M. K. Osentowski BA and D. W. Holt, “Evaluating the Efficacy of Intra-Aortic Balloon Pump Timing Using the Auto-Timing Mode of Operation With the Datascope CS100.”

J. T. Christenson, P. Badel, F. Simonet, and M. Schmuziger, “Preoperative Intra-aortic balloon pump enhances cardiac performance and improves the outcome of redo CABG,” Annals of Thoracic Surgery, vol. 64, no. 5, pp. 1237–1244, Nov. 1997, Doi: 10.1016/S0003-4975(97)00898-9.

H. Hausmann et al., “Prognosis after the implantation of an intra-aortic balloon pump in cardiac surgery calculated with a new score,” Circulation, vol. 106, no. 13 SUPPL., Sep. 2002, Doi: 10.1161/01.cir.0000032909.33237.f8.

S. R. Hosmane and A. G. Dawson, “In patients coming to theatre with an intra-aortic balloon pump, is it better to turn it off or keep it on while on bypass?” Interact Cardiovasc Thorac Surg, vol. 11, no. 3, pp. 314–321, Sep. 2010, Doi: 10.1510/icvts.2010.237255.

C. Kolyva, G. Biglino, J. R. Pepper, and A. W. Khir, “A Mock Circulatory System With Physiological Distribution of Terminal Resistance and Compliance: Application for Testing the Intra-Aortic Balloon Pump,” Artif Organs, vol. 36, no. 3, 2012, Doi: 10.1111/j.1525-1594.2010.01071.x.

A. Kucuker et al., “Single-centre experience with perioperative use of Intra-aortic balloon pump in cardiac surgery,” Heart Lung Circ, vol. 23, no. 5, pp. 475–481, 2014, Doi: 10.1016/j.hlc.2013.11.005.

R. Murthy P, H. S. N. Setty, G. Kamalapurkar, R. K. Nagashetty, and C. N. Manjunath, “Beneficial Effects of Pre-Operative Intra Aortic Balloon Pump Support in High Risk Patients Undergoing Coronary Artery Bypass Graft Surgery,” Heart Research - Open Journal, vol. 4, no. 2, pp. 23–28, Jul. 2017, Doi: 10.17140/HROJ-4-137.

F. Liu et al., “Timing of Intra-Aortic Balloon Pump Placement Before Off-Pump Coronary Artery Bypass Grafting and Clinical Outcomes,” Artif Organs, vol. 42, no. 3, pp. 263–270, Mar. 2018, Doi: 10.1111/aor.13009.

G. Gatti et al., “Preoperative Intra-Aortic Counterpulsation in Cardiac Surgery: Insights From a Retrospective Series of 588 Consecutive High-Risk Patients,” J Cardiothorac Vasc Anesth, vol. 32, no. 5, pp. 2077–2086, Oct. 2018, Doi: 10.1053/j.jvca.2017.12.008.

M. M. Townsley, “Prophylactic Intra-aortic Balloon Counterpulsation—Still Searching for Answers,” Journal of Cardiothoracic and Vascular Anesthesia, vol. 32, no. 5. W.B. Saunders, pp. 2074–2076, Oct. 01, 2018. Doi: 10.1053/j.jvca.2018.01.051.

E. Litton et al., “Six-Month Outcomes After High-Risk Coronary Artery Bypass Graft Surgery and Preoperative Intra-aortic Balloon Counterpulsation Use: An Inception Cohort Study,” J Cardiothorac Vasc Anesth, vol. 32, no. 5, pp. 2067–2073, Oct. 2018, Doi: 10.1053/j.jvca.2018.01.005.

K. Nakamura et al., “The use of prophylactic intra-aortic balloon pump in high-risk patients undergoing coronary artery bypass grafting,” PLoS One, vol. 14, no. 10, Oct. 2019, Doi: 10.1371/journal.pone.0224273.

G. Samanidis, G. Georgiopoulos, S. Bousounis, P. Zoumpourlis, and K. Perreas, “Outcomes after intra-aortic balloon pump insertion in cardiac surgery patients,” Rev Bras Ter Intensiva, vol. 32, no. 4, pp. 542–550, Oct. 2020, Doi: 10.5935/0103-507X.20200091.

S. Gelsomino et al., “Is visceral flow during intra-aortic balloon pumping size or volume dependent?,” Perfusion (United Kingdom), vol. 32, no. 4, pp. 285–295, May 2017, Doi: 10.1177/0267659116678058.

H. Parissis, V. Graham, S. Lampridis, M. Lau, G. Hooks, and P. C. Mhandu, “IABP: History-evolution-pathophysiology-indications: What we need to know,” Journal of Cardiothoracic Surgery, vol. 11, no. 1. BioMed Central Ltd., Aug. 04, 2016. Doi: 10.1186/s13019-016-0513-0.

S. Anand, T. Barry, R. Arsanjani, and L. LeMond, “Echocardiography in Cardiac Assist Devices,” Reviews in Cardiovascular Medicine, vol. 23, no. 7. IMR Press Limited, Jul. 01, 2022. Doi: 10.31083/j.rcm2307253.

R. Shiraishi, Y. Okazaki, K. Naito, and T. Itoh, “Perforation of the Descending Aorta by the Tip of an Intra-Aortic Balloon Pump Catheter.,” Circulation Journal, vol. 66, no. 4, pp. 423–424, 2002, Doi: 10.1253/circj.66.423.

J. T. Kim et al., “The carina as a useful radiographic landmark for positioning the Intra-aortic balloon pump,” Anesth Analg, vol. 105, no. 3, pp. 735–738, Sep. 2007, Doi: 10.1213/01.ane.0000278086.23266.35.

A. J. Rastan et al., “Visceral arterial compromise during intra-aortic balloon counterpulsation therapy,” Circulation, vol. 122, no. 11 SUPPL. 1, Sep. 2010, Doi: 10.1161/CIRCULATIONAHA.109.929810.

M. A. Klopman, E. P. Chen, and R. M. Sniecinski, “Positioning an Intra-aortic balloon pump using intraoperative transesophageal echocardiogram guidance,” Anesth Analg, vol. 113, no. 1, pp. 40–43, 2011, Doi: 10.1213/ANE.0b013e3182140b9a.

H. Parissis, A. Soo, M. Leotsinidis, and D. Dougenis, “A statistical model that predicts the length from the left subclavian artery to the celiac axis; towards accurate intra aortic balloon sizing,” J Cardiothorac Surg, vol. 6, no. 1, Aug. 2011, Doi: 10.1186/1749-8090-6-95.

V. Venugopal, “A Novel Technique for Intra-aortic Balloon Positioning in the Intensive Care Unit,” J Extra Corpor Technol. vol.44, no. 3, pp. 160-162, Sep. 2012, PMID: 23198398; PMCID: PMC4557529.

Downloads

Published

31-10-2023

How to Cite

Analysis of Intra-Aortic Balloon Performance in Open-Heart Surgery. (2023). Al-Nahrain Journal for Engineering Sciences, 26(3), 157-166. https://doi.org/10.29194/NJES.26030157

Similar Articles

1-10 of 35

You may also start an advanced similarity search for this article.