Mechanical Analysis of Bone-Plate Construct Regarding Strength and Stiffness

Abstract

The aim of this study was to support surgeons to decide where to place the screws in order to achieve an optimal fracture healing and to prevent implant failure after a femoral shaft fracture So this paper focus on the analysis of bone-plate construct by using Finite element Analysis (FEA), comminuted femur fractured bone fixed with Dynamic Compression Plate (DCP) 16 holes by 4.5 Cortex screws, to investigate the effects of screws configuration on the mechanical behavior of different seven model as Interfragmentary strain which is the most important factor for femur fracture healing. The results state the relationships between the Von-Mises stress, Total deformation and Interfragmentary strain with respect to the screws configuration. The study shows the regions of maximum stress from stress distribution and also founded that we can decrease the Interfragmentary strain by increasing the number of screws.

References

  1. M. Gudarzi, H. Zamanian, B. Marzban, and S. Amoozegar, “Investigation of the Stress Distribution in a Bone Due to Screws Pretensions of the Bone Plate,” Elastic, vol. 20, pp. 0.3, 2012.
  2. M. Schimmelpfenning. "Patient specific optimization of fracture treatment considering the inhomogeneous material properties of bone tissue and the expected load situation," https://www.ozeninc.com/wpcontent/uploads/2014/12/03_WOST2013_Optimization_Schimmelpfennig_Paper.pdf.
  3. G. R. Kristina Haase, “Finite Element Analysis of Orthopaedic Plates and Screws to Reduce the Effects of Stress Shielding,” Master of Applied Science in Mechanical Engineering University of Ottawa, Ontario 2009.
  4. B. Wongchai, “The Effect of the Configuration of the Screw Fixation on the Interfragmentary Strain,” American Journal of Applied Sciences, vol. 9, pp. 842-845, 1/6/2012.
  5. J. K. Oh, D. Sahu, Y. H. Ahn, S. J. Lee, S. Tsutsumi, J. H. Hwang, et al., "Effect of fracture gap on stability of compression plate fixation: a finite element study," J Orthop Res, vol. 28, pp. 462-7, Apr 2010
  6. P. S. R. S. Maharaj, R. Maheswaran, and A. Vasanthanathan, "Numerical Analysis of Fractured Femur Bone with Prosthetic Bone Plates," Procedia Engineering, vol. 64, pp. 1242-1251, 2013.
  7. U. Mughal, H. Khawaja, and M. Moatamedi, “Finite element analysis of human femur bone,” The International Journal of Multiphysics, vol. 9, no. 2, 2016.
  8. D. Amalraju, and A. S. Dawood, “Mechanical strength evaluation analysis of stainless steel and titanium locking plate for femur bone fracture,” Engineering Science and Technology: An International Journal, vol. 2, no. 3, pp. 381-388, 2012.
  9. S. Das, and S. K. Sarangi, “Finite element analysis of femur fracture fixation plates,” International Journal of Basic and Applied Biology, vol. 1, no. 1, pp. 1-5, 2014.
  10. P. K. Satapathy, B. Sahoo, L. N. Panda, and S. Das, "Finite element analysis of functionally graded bone plate at femur bone fracture site," IOP Conference Series: Materials Science and Engineering, vol. 330, p. 012027, 2018.